The exposed mechanical and electronic components inside a CT gantry.
A view inside a CT scanner shows the hardware surrounding its circular patient opening. Nettadi / Wikimedia Commons · CC BY 4.0. Image credit & source ↗

A study published on 28 July 2026 updated U.S. benchmarks for common adult CT examinations using data from 2025. Across 5.23 million acquisitions at 592 facilities, the authors found lower national CTDIvol achievable-dose and diagnostic-reference values than in the 2014 comparison. These are population-level tools for protocol review: they are not measurements of an individual patient's absorbed dose, and a lower benchmark does not by itself prove better imaging. [1] [3]

A benchmark starts with a defined measurement

CT scanners report quantities such as CTDIvol and dose-length product (DLP). CTDIvol is based on measurements in a standardized cylindrical phantom and is useful for comparing scanner output under defined conditions. DLP combines that index with scan length. Neither number is a direct measurement of the energy absorbed throughout a particular patient's body. Patient size and the scanner's output both matter; size-specific dose estimate (SSDE) uses a size correction to make an estimate more relevant to the person being scanned. [3]

A diagnostic reference level (DRL) answers a different question: how does a facility's typical dose for a defined examination compare with a broader distribution? In this study, the authors calculated facility-level median doses, then summarized their distributions. The 50th percentile was the achievable dose (AD), and the 75th percentile was the DRL. These statistics describe practice across facilities; they are not individual-patient limits or targets that every scan must meet. [1]

A 2025 snapshot, published in 2026

The research team analyzed 5,234,285 CT acquisitions from 2,802,416 adults at 592 U.S. facilities. The scans were performed from January through December 2025 and covered ten common examination categories, matched to categories used in the 2014 benchmark study. The paper appeared in Radiology on 28 July 2026. That distinction matters: 2025 is the period represented by the measurements, while July 2026 is when the study became a published result. [1]

The study used standardized examination names and dose reports, estimated patient size from water-equivalent diameter, and calculated SSDE with established medical-physics methods. This makes comparisons more informative than treating every exam as interchangeable. A head CT, a chest CT and a longer combined examination do not have the same scan geometry or clinical task, so dose statistics need to remain tied to the examination category and size assumptions that produced them. [1] [3]

What the percentiles say

Compared with the 2014 benchmark, the study reported a 9.3% decrease in CTDIvol achievable doses and a 21.8% decrease in CTDIvol DRLs across the national adult comparisons. The largest reported reduction was for chest examinations with contrast material, at 31.2%; the smallest was for head examinations without contrast, at 3.5%. The paper also reported that dose measures remained dependent on patient size for torso examinations. These are changes in benchmark statistics, not a claim that every facility or every patient's dose fell by those percentages. [1]

A small hypothetical example shows why the two percentiles are different. Suppose the median dose at each of several facilities is first calculated for the same examination category. The 50th percentile of that facility distribution marks its midpoint; the 75th percentile marks a higher point, with one quarter of facility medians above it. If both values shift downward in a later benchmark, the typical and upper-range facility practices have changed by different amounts. No single person's scan can be located from those national summaries alone. [1]

Size normalization changes the comparison

An output index can appear different when people of different sizes are mixed together. In body CT, the study used water-equivalent diameter, which incorporates how strongly the patient's cross-section attenuates X-rays, and calculated SSDE to account for size. AAPM guidance explains why this correction matters: scanner output indices are referenced to phantoms and do not themselves estimate patient dose. SSDE is a more size-aware estimate, but it still is not a complete map of absorbed dose to every organ. [1] [3]

A fair comparison also needs similar examination definitions, scan coverage and dose quantities. Otherwise, a changed mix of patients or protocols can look like a change in scanner performance. Size normalization improves the comparison; it does not convert a benchmark into an individualized dose calculation. [1] [3]

Useful for quality review, not a treatment rule

A facility can use a DRL as a signal to review an examination category whose typical output is high relative to peers. The American College of Radiology describes its Dose Index Registry as a way to compare facility data with broader and similar-facility groups, identify outliers, and support optimization while maintaining image quality. A flagged value starts a technical review: staff can check protocol, patient size, scan length, number of phases and whether the images answer the clinical question. It does not automatically mean that a specific scan was inappropriate. [2]

The 2026 paper is a large retrospective benchmarking study, not a randomized test of a dose-reduction intervention. Its scale and updated measurements make the comparison useful, but the participating facilities and included examination categories define what it can represent. The results do not establish why doses changed, whether every patient's exposure changed, or whether any one image was diagnostically sufficient. Clinical justification and image quality remain part of the evaluation. [1] [2]

The advance is a newer, size-aware reference for comparing adult CT practice across facilities. Keep the measurement chain in view: scanner output, size adjustment, facility distributions and clinical task are related but distinct. Benchmarks can guide system review; they cannot replace an individual's medical assessment or decide whether a scan is needed. [1] [3]

Explore the physics: an explanatory diagram
A CT source and opposing curved detector array rotate around a schematic body cross-section; multiple projections form reconstructed slices.
CT collects X-ray transmission measurements from many directions using a rotating source and opposing detectors. A computer reconstructs cross-sectional images from the measurements. The body contours, projections, and slice stack are conceptual, not patient data. Illustration credit ↗
FOLLOW THE EVIDENCE

Sources & further reading

  1. A Decade of Change in Adult CT Radiation Doses: 2025 U.S. Diagnostic Reference Levels Radiology / Radiological Society of North America
  2. DIR Computed Tomography Features American College of Radiology
  3. Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations (AAPM Report No. 204) American Association of Physicists in Medicine

Sources checked: 28 November 2026. The calculations illustrate physical models; they are not clinical settings.

Prepared for the independent PSI Center reading library. About edition dates and source checking.

This explainer describes scientific principles. It does not assess individual treatment choices.